TY - JOUR
T1 - Microdevice for continuous flow magnetic separation for bioengineering applications
AU - Khashan, Saud Abdelaziz
AU - Dagher, Sawsan
AU - Alazzam, Anas
AU - Mathew, Bobby
AU - Al Naqbi, Ali Abdullah Hilal
N1 - Funding Information:
The Authors acknowledge financial support received from the UAEU Start-up research fund, project no. 31N158, UAE.
Publisher Copyright:
© 2017 IOP Publishing Ltd.
PY - 2017/4/7
Y1 - 2017/4/7
N2 - A novel continuous flow microfluidic device, integrated with soft-magnetic wire (permalloy), is fabricated and tested for magnetophoresis based separation. The flow-invasive permalloy wire, magnetized using an external bias field, is positioned perpendicular to the external magnetic field and with its length traversing the introduced sample flow. The microfluidic device is realized in PDMS; the mold for PDMS microstructures is cut out of Plexiglas® sheets with controllable dimensions. Microfluidic devices with microchannel height ranging between 0.5 mm and 2 mm are fabricated. Experiments are carried out with and without sheath flow; with sheath flow the microparticles are focused at the center of the microchannel. When focusing is not employed, the microdevice can exhibit a complete separation (or filtration) with the introduction of the sample at rates lower than a maximum threshold. However, this complete separation is attributed to the fact that part of the particles, once they approach the repulsive field of the wire, will find their way into the attractive region of the wire while the remaining will be indefinitely trapped at the channel walls. On the other hand, when the focused sample is flowing at the same rate but alongside an appropriate sheath flow, the complete separation can be achieved with all (initially repelled) particles being captured on the attractive region of the wire itself. This microdevice design is well suited for purification, enrichment, and detection of microparticles in lab-on-a-chip devices due to its ability to handle high throughput without compromising capture efficiency while exhibiting excellent reliability and flexibility.
AB - A novel continuous flow microfluidic device, integrated with soft-magnetic wire (permalloy), is fabricated and tested for magnetophoresis based separation. The flow-invasive permalloy wire, magnetized using an external bias field, is positioned perpendicular to the external magnetic field and with its length traversing the introduced sample flow. The microfluidic device is realized in PDMS; the mold for PDMS microstructures is cut out of Plexiglas® sheets with controllable dimensions. Microfluidic devices with microchannel height ranging between 0.5 mm and 2 mm are fabricated. Experiments are carried out with and without sheath flow; with sheath flow the microparticles are focused at the center of the microchannel. When focusing is not employed, the microdevice can exhibit a complete separation (or filtration) with the introduction of the sample at rates lower than a maximum threshold. However, this complete separation is attributed to the fact that part of the particles, once they approach the repulsive field of the wire, will find their way into the attractive region of the wire while the remaining will be indefinitely trapped at the channel walls. On the other hand, when the focused sample is flowing at the same rate but alongside an appropriate sheath flow, the complete separation can be achieved with all (initially repelled) particles being captured on the attractive region of the wire itself. This microdevice design is well suited for purification, enrichment, and detection of microparticles in lab-on-a-chip devices due to its ability to handle high throughput without compromising capture efficiency while exhibiting excellent reliability and flexibility.
KW - biofluid
KW - continuous particle separation
KW - lab-on-a-chip
KW - magnetophoresis
KW - microfabrication
KW - microfluidics
UR - http://www.scopus.com/inward/record.url?scp=85018506061&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85018506061&partnerID=8YFLogxK
U2 - 10.1088/1361-6439/aa666d
DO - 10.1088/1361-6439/aa666d
M3 - Article
AN - SCOPUS:85018506061
SN - 0960-1317
VL - 27
JO - Journal of Micromechanics and Microengineering
JF - Journal of Micromechanics and Microengineering
IS - 5
M1 - 055016
ER -