T1 - Monitoring quality indicators for the Xpert MTB/RIF molecular assay in Ethiopia
AU - Alemu, Ayinalem
AU - Amare, Misikr
AU - Ameni, Gobena
AU - Kebede, Abebaw
AU - Beyene, Dereje
AU - Yenew, Bazezew
AU - Diriba, Getu
AU - Mehamd, Zemedu
N1 - Publisher Copyright:
© 2019 Kebede et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
PY - 2019/11/1
Y1 - 2019/11/1
N2 - Introduction In Ethiopia, >300 GeneXpert instruments have been deployed for tuberculosis (TB) testing using the Xpert MTB/RIF cartridge. Implementing quality indicators is necessary for monitoring and evaluating the quality of Xpert MTB/RIF diagnostic services. Objective To assess the use of quality indicators for the Xpert MTB/RIF molecular assay in Ethiopia and to compare the findings with the predefined targets described in the literature. Methods Clinical specimens collected from patients with suspected TB were subjected to Xpert MTB/ RIF testing at the National TB Reference Laboratory (NTRL) between January and December 2018. Data were collected from GeneXpert software and Laboratory Information System (LIS) databases. Quality indicators were calculated and analyzed. Bivariate and multivariate analyses were performed using SPSS software version 20 (SPSS Inc., Chicago, Illinois, USA). Results Of the 2515 specimens tested, 2274 (90.4%) had successful test results; 18.2% were positive for Mycobacterium tuberculosis (MTB). Among MTB positives (n = 413), 4.8% and 1.0% were rifampicin (RIF)-resistant and RIF-indeterminate cases, respectively. Unsuccessful results were 241 (9.6%); 8.9% of the total number of tests were errors, 0.04% had invalid results and 0.6% ‘no result’. The most frequent error was probe check failure (error 5007). Instrument module A4, B2, B3, C3, and D3 (p<0.05) and tester experience (p<0.05) had a statistically significant association with errors in multivariate analysis. Additional 42 MTB cases (9.2% of the total cases) were detected among unsuccessful results by follow-up tests. Sixty-four percent of the initial test results were released within the turnaround time (TAT) ≤24 hours. Conclusion Most of the quality indicators for the Xpert MTB/RIF molecular assay were maintained within the targets. However, the error rate and TAT were out of the targets. Defective modules and lacking experience were the factors affecting successful test outcomes.
AB - Introduction In Ethiopia, >300 GeneXpert instruments have been deployed for tuberculosis (TB) testing using the Xpert MTB/RIF cartridge. Implementing quality indicators is necessary for monitoring and evaluating the quality of Xpert MTB/RIF diagnostic services. Objective To assess the use of quality indicators for the Xpert MTB/RIF molecular assay in Ethiopia and to compare the findings with the predefined targets described in the literature. Methods Clinical specimens collected from patients with suspected TB were subjected to Xpert MTB/ RIF testing at the National TB Reference Laboratory (NTRL) between January and December 2018. Data were collected from GeneXpert software and Laboratory Information System (LIS) databases. Quality indicators were calculated and analyzed. Bivariate and multivariate analyses were performed using SPSS software version 20 (SPSS Inc., Chicago, Illinois, USA). Results Of the 2515 specimens tested, 2274 (90.4%) had successful test results; 18.2% were positive for Mycobacterium tuberculosis (MTB). Among MTB positives (n = 413), 4.8% and 1.0% were rifampicin (RIF)-resistant and RIF-indeterminate cases, respectively. Unsuccessful results were 241 (9.6%); 8.9% of the total number of tests were errors, 0.04% had invalid results and 0.6% ‘no result’. The most frequent error was probe check failure (error 5007). Instrument module A4, B2, B3, C3, and D3 (p<0.05) and tester experience (p<0.05) had a statistically significant association with errors in multivariate analysis. Additional 42 MTB cases (9.2% of the total cases) were detected among unsuccessful results by follow-up tests. Sixty-four percent of the initial test results were released within the turnaround time (TAT) ≤24 hours. Conclusion Most of the quality indicators for the Xpert MTB/RIF molecular assay were maintained within the targets. However, the error rate and TAT were out of the targets. Defective modules and lacking experience were the factors affecting successful test outcomes.
UR - http://www.scopus.com/inward/record.url?scp=85074878550&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85074878550&partnerID=8YFLogxK
U2 - 10.1371/journal.pone.0225205
DO - 10.1371/journal.pone.0225205
M3 - Article
C2 - 31714934
AN - SCOPUS:85074878550
SN - 1932-6203
VL - 14
JO - PLoS ONE
JF - PLoS ONE
IS - 11
M1 - e0225205
ER -